Elimko TC01-1S5I22-100-PZ is a straight-type single-element type S (Pt%10Rh-Pt (IEC 60584)) platinum thermocouple, Ø22 mm, with 1000 mm immersion length. Impact-resistant Inconel metal tube outside, KER 799 ceramic tube inside. Used below 1200 °C where a bare ceramic tube could break and where reference-grade accuracy is required. This is the example code in the Elimko catalogue.
Platinum thermocouples are far more stable than type K above 1200 °C, so they are used in ceramic, glass, sintering and high-temperature heat-treatment furnaces. The platinum element is sensitive to metal vapours and reducing atmospheres; the right protection tube determines its life.
Technical data
| Order code | TC01-1S5I22-100-PZ |
|---|---|
| Design | TC01 – straight type with metal protection tube |
| Element | Single, type S, Pt%10Rh-Pt (IEC 60584), wire diameter 0.5 mm |
| Measuring range (type) | 0 … 1500 °C |
| Protection tube | Inconel metal protection tube Ø22 mm |
| Tube outer diameter | 22 mm |
| Inner tube | KER 799 pure alumina inner tube (PZ) |
| Immersion length | 1000 mm |
| Temperature limits | Element: 1300 °C continuous, 1600 °C short-term (Ø0.5 mm wire) · Tube limit: 1200 °C |
| Usable upper limit | 1200 °C (lower of tube and element limits, short-term) |
| Connection head | Aluminium head (A/B/C); head temperature max. 200 °C |
| Extension cable | Use the matching S/R compensating cable |
Order code structure: TC0x – number of elements + type (S/R/B) + wire diameter + tube material + tube diameter – length (cm) – inner tube (PY: KER 610, PZ: KER 799). Tube material codes: X = KER 530, Y = KER 610, Z = KER 799, I = Inconel. Ask for type B (Pt30%Rh-Pt6%Rh), dual elements, flanges and calibration certificates.
Technical data is based on the Elimko catalogue (Straight Type Thermocouples, Thermocouple General Information) and is for information only. Elimko thermocouples are made to order; we define the order code together at quotation stage for other lengths, diameters, protection tubes, inner tubes, flanges and heads. Temperature limits apply in air without harmful gases; they are lower in reducing or corrosive atmospheres.
For price and lead time, use the Request a Quote button or call +90 533 945 27 94.
Tags: Elimko TC01-1S5I22-100-PZ, type S thermocouple, platinum thermocouple, ceramic tube thermocouple, high temperature thermocouple, 1000 mm, Elimko